Calibration

Unboxing BLZ Lab Scanner LS100

What are the computer specifications for Lab Scanner LS100?

Scanner installation

Calibration

Calibration

When will you need to calibrate?

  • Use the scanner for the first time;
  • The scanner has not been calibrated for a long time (the scanning software will automatically remind you of the calibration every seven days);
  • The scanner has been shaken, vibrated or impacted. The scanned model data is abnormal, the surface is rough and does not match the actual object.
Calibration procedure

1. Open the calibration software.

2. Click “Calibrate” in the settings page.

3. Place the calibration tool on the heightening mat.

4. Click  “Start” and the scanner will start to calibrate automatically.

5. Completed.